Model 4500 Premier Dust and Opacity Monitor
61
Protocol Format
The Model 4500 Premier utilises the MODBUS RTU (Remote Terminal Unit) transmission format.
Serial communications are fixed at 1 start bit, 8 data bits, no parity, 1 stop bit, and the data rate is
fixed at 9600 baud. The following are examples of master query and slave response messages in
RTU format, polling slave address 11 (hex).
(03) Read Register(s)
Query
Response
Slave Address
11
Slave Address
11
Function
03
Function
03
Register Address Hi
00
Byte Count
06
Register Address Lo
20
Data 1 Hi
xx
No. of Registers Hi
00
Data 1 Lo
xx
No. of Registers Lo
03
Data 2 Hi
xx
CRC Hi
xx
Data 2 Lo
xx
CRC Lo
xx
Data 3 Hi
xx
Data 3 Lo
xx
CRC Hi
xx
CRC Lo
xx
Calibration Phases (Register 168)
Three bits in this register are defined: all other bits are zero. If one bit is true then the others are false.
Bit 0 Zero Phase
True during a Calibration Check sequence.
Goes true Cal Action delay seconds before the zero reflector starts to drive in.
Goes false before the upscale filter starts to drive in.
Bit 1 Span Phase
True during Calibration Check sequence.
Goes true Cal Action Delay seconds before the upscale filter starts to drive in (ie simultaneously with
Bit 0 going false)
Goes false Cal Action Delay seconds before the zero reflector and upscale filter start to drive out.
Bit 2 Audit Phase
True throughout any Calibration Audit operation.
Goes true Cal Action Delay seconds before the zero reflector starts to drive in.
Goes false Cal Action Delay seconds before the zero reflector starts to drive out.