Model 4500 Premier Dust and Opacity Monitor
19
A3-6 Using the Flange Alignment Tool
The flange alignment tool (part number 702.787) consists of two plates which are mounted on to the
flanges. One plate supports a light source, the other a small telescope with a target in the centre. When
the flanges are properly aligned, the light spot is located in the centre of this target.
The flange alignment tool is used as follows:
1
Remove the 12 washers from the mounting flanges, retain these as they are essential to the
correct installation of the instrument.
2
Tack weld the flanges in position.
3
Mount the flange alignment tool as shown above.
4
Align flange number 1 and weld it into place.
5
Reverse the flange alignment tool (switch components).
6
Align flange number 2 and weld it into place.
7
Replace the washers, see section A3-7 installation details.
Flange 1
Flange 2
Light Source
Light Source
Alignment Circle
reinforcement if
required