Model 4500 Premier Dust and Opacity Monitor
102
MANUAL CALIBRATION (MANUAL CAL)
This sequence is not recommended to be followed by the user. The manual
cal sequence requires values to be entered manually. Re-calibrate is a
more accurate operation. This provision is of more practical use to a
service engineer.
FACTORY SETTINGS (RECALL FACTORY)
Provides a simple method to return the Model 4500 Premier to its factory
calibration state. This may be helpful e.g. where a re-calibrate sequence
has not been successful. On selection of ‘Recall Factory’ from the
Calibration menu press the
‘
ENTER' (
↵
↵↵
↵↵
) key. The display reads ‘Restore
Factory’, press the
‘
ENTER' (
↵
↵↵
↵↵
) key again and then select YES or NO
using the
←
or
→
key.
Press the
‘
ENTER' (
↵
↵↵
↵↵
) key whilst ‘YES’ is displayed to return the
instrument back to its factory settings. Press the
‘
ENTER' (
↵
↵↵
↵↵
) key whilst
‘NO’ is displayed to return to the ‘Recall Factory’ screen on the
Calibration menu.
CALIBRATION SETUP (CAL SETUP)
On selection of the ‘Cal. Setup’ sub-menu press the ‘ENTER’ (
↵
↵↵
↵↵
) key. To
set the time interval between calibrations (Interval x.xxh) and the time
that the instrument will wait before performing the first calibration (Time
to next cal). These two options appear as a sub-menu and are scrolled
by pressing the
↑
or
↓
keys.
↵
↵↵
↵↵
↵
↵↵
↵↵
↵
↵↵
↵↵
→
←
↵
↵↵
↵↵
Opacity
0%
Manual Cal.
Opacity
0%
Restore Factory:
Yes
Opacity
0%
Restore Factory:
No
Opacity
0%
Restore Factory
Opacity
0%
Recall Factory
Opacity
0%
Factory Values
Opacity
0%
Cal. Setup
Interval
↑↓
Time to next cal.