4 Maintenance
34
QuickGene-810 Operation Manual
Rev. 1.7
■
How to Replace the Pump Tube
Perform the following procedures to replace the Pump Tube.
(1) Turn the power on.
Press the [I] side of the switch.
(2) Press the [DISCHARGE] button.
All the liquid in the tubes will be discharged.
In addition, pressing the [DISCHARGE] button
makes the pump rotate to the position for attaching
the pump cover.
(3) Turn off the power.
Press the [O] side of the switch.
(4) Open the front cover.
Hold the handle and open the front cover until the
upper limit is reached.
(5) Take off both the tube on the device side
and the Tube Cap from the Pump Tube.
Take off both tube caps (purple and blue) by turning
the connection points as shown in the figure.
(6) Take off the Buffer Stand from the
QuickGene-810.
(7) Detach the pump cover.
Loosen and remove the screws to detach the pump
cover.
CAUTION:
The front cover will not stay
on the way to the upper
limit. If you release the front
cover while opening it, it
may cause injury or device
damage.
Tube on the device side
Tube cap
Summary of Contents for QuickGene-810
Page 1: ...Rev 1 7 Original Instructions Operation Manual Rev 2 1 ...
Page 2: ......
Page 6: ...1 2 6 3 4 5 7 iv QuickGene 810 Operation Manual Rev 1 7 ...
Page 14: ...1 2 6 3 4 5 7 xii QuickGene 810 Operation Manual Rev 1 7 KURABO INDUSTRIES ...
Page 19: ...QuickGene 810 Operation Manual Rev 1 7 xvii 1 3 2 5 4 6 7 ...
Page 20: ...1 2 6 3 4 5 7 xviii QuickGene 810 Operation Manual Rev 1 7 ...
Page 24: ...1 2 6 3 4 5 7 xxii QuickGene 810 Operation Manual Rev 1 7 ...
Page 60: ...1 2 6 3 4 5 7 4 Maintenance 36 QuickGene 810 Operation Manual Rev 1 7 ...
Page 76: ......