About Installation Conditions
For correct and safe use of QuickGene-810, install it in a location that meets the following conditions:
• A location where power can be provided. Use an AC100±10% to AC240V±10%, 50Hz or
60Hz outlet.
• A location where the temperature is 15 to 30°C, and humidity is 30 to 80%RH (non
condensing).
• A location that will support the weight of QuickGene-810 (21kg), that is flat and stable, with no
vibration.
• A location away from direct sunlight. Block the sunlight by curtains or blinds as necessary.
• A location which is well-ventilated and not dusty.
• A location where the temperature does not go up and down suddenly. Warming a cold room
suddenly or moving the QuickGene-810 from a room with low temperature to a warm room
may cause condensation inside the device (condensing), and may have a negative influence
on the sample result.
• A location far from water taps, water heaters, humidifiers, air-conditioners and heaters. (Avoid
areas of high temperature and high humidity, or low temperature and low humidity)
• A location far from objects which generate strong magnetic fields (motors, transformers, TV,
audio speakers, magnets, etc.). Bringing the QuickGene-810 close to any type of magnetic
field may cause a malfunction.
Product Use Limitations
•
FUJIFILM Corporation
assumes no liability for results regardless of the intended
use.
• It is the user's responsibility to validate the performance of QuickGene kits and
QuickGene-810 for any case of use.
• The QuickGene-810 is a device for research and study only.
When you use the QuickGene-810, read this manual thoroughly and follow the safety management
regulations of your facilities (laboratory).
WARNING:
•
Do not install the QuickGene-810 to a location where it can be
splashed with water. Doing so may cause a device failure, fire, or
electric shock.
•
When relocating the QuickGene-810, be sure to disconnect the plug
from the outlet first. Watch your step when you relocate the
QuickGene-810. If the power cable is damaged, this may cause a
device failure, fire, injury, or electric shock.
CAUTION:
•
Do not install the QuickGene-810 in an unstable place such as a
slanted surface or a place subject to vibrations. Doing so may cause
injury or device failure.
•
Do not install the QuickGene-810 in direct sunlight or close to a
heating device. Doing so may shorten the life of QuickGene-810, or
cause a failure.
•
Two or more people are necessary to relocate the QuickGene-810.
Otherwise, you may hurt your back or hands because of the weight of
QuickGene-810 (more than 20kg).
KURABO INDUSTRIES
Summary of Contents for QuickGene-810
Page 1: ...Rev 1 7 Original Instructions Operation Manual Rev 2 1 ...
Page 2: ......
Page 6: ...1 2 6 3 4 5 7 iv QuickGene 810 Operation Manual Rev 1 7 ...
Page 14: ...1 2 6 3 4 5 7 xii QuickGene 810 Operation Manual Rev 1 7 KURABO INDUSTRIES ...
Page 19: ...QuickGene 810 Operation Manual Rev 1 7 xvii 1 3 2 5 4 6 7 ...
Page 20: ...1 2 6 3 4 5 7 xviii QuickGene 810 Operation Manual Rev 1 7 ...
Page 24: ...1 2 6 3 4 5 7 xxii QuickGene 810 Operation Manual Rev 1 7 ...
Page 60: ...1 2 6 3 4 5 7 4 Maintenance 36 QuickGene 810 Operation Manual Rev 1 7 ...
Page 76: ......