QuickGene-810 Operation Manual
Rev. 1.7
37
5 How to Set Parameters
1
3
2
5
4
6
7
5
How to Set Parameters
This chapter explains how to set parameters.
"Set Parameters" means to change the values (such as buffer volume, number of discharges, etc.) for each
extraction mode.
The name of each extraction mode can be selected from 35 names.
The parameters set in the QuickGene-810 are stored in 7 separate memory areas. Therefore, even if an
extraction mode name is changed, the parameter values will not be changed.
•
"5.2 How to Set/Change Parameters" (page 39)
•
"5.4 Changing Extraction Mode Names" (page 43)
5.1 Parameters
Settings of each parameter can be changed from the setup mode.
Some parameters are changed from the normal setup mode (USER mode) and others are changed from the
EXPERT mode.
The following table shows the display sequence, setting range, default value, etc. for each parameter in
USER mode.
Important: • The setup is used to change the parameters, according to new protocol of
the kit, etc. Please check the parameters for each kit's Handbook.
• In the case of wrong value for parameters, it will cause buffer spill or
sample extraction failure.
It is the user's responsibility to change the parameters.
• In EXPERT mode, there are some crucial parameters. Therefore, you must
not change any parameters in EXPERT mode without our guidance.
Display
sequence
LCD message
Setting
range
Unit
Mode name (The value displays the default)
DNA
WHOLE
BLOOD
RNA
CELL
RNA
CELL
PLUS
DNA
TISSUE
RNA
TISSUE
RNA
TISSUE
PLUS
PLASMID
1 BIND PEAK
50 - 160
kpa
120
120
120
120
120
120
120
2 WASH COUNT
0 - 5
time(s)
3
3
1
3
3
1
2
3 WASH PEAK
50 - 160
kpa
110
110
110
110
110
110
110
4 WASH VOL1
250 - 750
µl
750
500
500
750
750
750
750
5 WASH VOL2
250 - 750
µl
750
500
500
750
750
750
750
6 WASH VOL3
250 - 750
µl
750
500
500
750
750
750
750
7 WASH VOL4
250 - 750
µl
750
500
500
750
750
750
750
8 WASH VOL5
250 - 750
µl
750
500
500
750
750
750
750
9 WASH DIP TM
0 - 510
sec
0
150
150
0
150
150
0
10 WAS2 WAIT T
0 - 255
min
0
0
5
0
0
5
0
11 WAS2 COUNT
0 - 5
time(s)
0
0
2
0
0
2
0
12 WAS2 PEAK
50 - 160
kpa
110
110
110
110
110
110
110
13 WAS2 VOL1
250 - 750
µl
750
500
500
750
750
750
750
14 WAS2 VOL2
250 - 750
µl
750
500
500
750
750
750
750
15 WAS2 VOL3
250 - 750
µl
750
500
500
750
750
750
750
16 WAS2 VOL4
250 - 750
µl
750
500
500
750
750
750
750
17 WAS2 VOL5
250 - 750
µl
750
500
500
750
750
750
750
18 ELUT VOL
50 - 400
µl
200
100
100
200
100
100
50
19 ELUT PEAK
50 - 160
kpa
100
100
100
100
100
100
100
20 ELUT DIP TM
0 - 510
time(s)
0
30
30
90
30
30
0
Summary of Contents for QuickGene-810
Page 1: ...Rev 1 7 Original Instructions Operation Manual Rev 2 1 ...
Page 2: ......
Page 6: ...1 2 6 3 4 5 7 iv QuickGene 810 Operation Manual Rev 1 7 ...
Page 14: ...1 2 6 3 4 5 7 xii QuickGene 810 Operation Manual Rev 1 7 KURABO INDUSTRIES ...
Page 19: ...QuickGene 810 Operation Manual Rev 1 7 xvii 1 3 2 5 4 6 7 ...
Page 20: ...1 2 6 3 4 5 7 xviii QuickGene 810 Operation Manual Rev 1 7 ...
Page 24: ...1 2 6 3 4 5 7 xxii QuickGene 810 Operation Manual Rev 1 7 ...
Page 60: ...1 2 6 3 4 5 7 4 Maintenance 36 QuickGene 810 Operation Manual Rev 1 7 ...
Page 76: ......