Portable spectrum analyzer Arinst SSA-TG R2 6 April 2019
www.arinst.net | www.arinst.ru | www.kroks.ru
50
igure 72 Selection of the reflection coefficient measurement mode
9.3.4. To reduce the influence of connecting wires and connectors on the measurement results, it is
necessary to calibrate.
Note. It is allowed to perform calibration once for modes
S11
and
SWR
, provided that the connectors
and cables of the test circuit do not change.
Note. To measure the reflection coefficient
S11
, you must use a directional coupler or measuring
bridge (not included). In our example in figure 19, the
KROKS KSB 2700
measuring bridge is specifically
designed for use with ARINST spectrum analyzers.
9.3.5. Connect the output of the signal generator (1)
TG OUT
of the device to the input of the measur-
ing bridge
TO TG OUT
according to the wiring diagram in figure 19.
Output of the reflected signal from the
TO RF IN
connect to the antenna input of the device (2)
RF IN
.
9.3.6. Connect loads from a standard set of calibration loads in series to the
TEST
of the measuring
bridge (not included in the package):
- Open standart and press the
Open calibration
on the screen;
- Short standart, and press the
Short
calibration
on the screen;
- Load standart click
Load calibration
.
After calibration commands
Open calibration
,
Short calibration
and
Load calibration
will be
changed to cancel commands:
Reset open calibration
;
Reset short calibration
and
Reset load cali-
bration
, figure 73.