Fct. 3.6.2 P-SENSOR
pressure sensor
Select
YES
if pressure sensor option is present else
select
NO.
LIMITS
-
Not applicable.
APPEARS - If fluid is not
LIQUID.
Fct. 3.6.3 RS485 ID
RS485 identifier
Use this menu to select whether identifier of the meter
should be used in RS485 transmission. If
YES
is
selected, then identifier number for the meter can be
entered using
Fct. 3.6.4 METER ID
menu.
LIMITS - Not applicable
APPEARS - If output type of the meter is selected as
RS 485.
Fct. 3.6.4 METER ID
Meter identifier
Use this menu to set the RS485 identifier number for
the meter. This number is used in RS485 transmission
and is transmitted with every data frame if
Fct. 3.6.3
RS485 ID
is selected as
YES.
LIMITS - 0 to 99
APPEARS - If output type of the meter is selected as
RS 485
and
Fct. 3.6.3 RS485 ID
is selected as
YES.
Fct. 3.7.0 P-DATA
This submenu groups functions which input the
pressure sensor calibration data. Pressure sensor
calibration data (mV output vs. pressure) is obtained
at two temperatures. So, it is possible to compensate
pressure sensor for temperature variations.
APPEARS - If
FLUID
is not
LIQUID
and
Fct.
3.6.2 P-SENSOR
is
YES.
Fct. 3.7.1 T1
calibration temperature 1
Enter the calibration temperature 1. Units possible to
use are same as in
Fct. 3.3.1.
LIMITS - 20 to 100 Deg. C.
APPEARS - if Fct. 3.6.2
P-SENSOR
is
YES
and
FLUID
is not
LIQUID
.
Fct. 3.7.2 P11 V11
calibration point 1 at T1
Enter pressure value (in one of the units same as in
Fct. 3.3.3
) and the corresponding milli-volt value.
LIMITS - 0 to 40 Bar for P11.
0 to 25 mV for V11.
APPEARS - if
Fct. 3.6.2 P-SENSOR
is
YES
and
fluid is not
LIQUID.
Fct. 3.7.3 P12 V12
calibration point 2 at T1
Enter pressure value (same unit assumed as in
Fct.
3.7.2
) and the corresponding milli-volt value.
LIMITS - 5 to 100 Bar for P12.
40 to 125 mV for V12.
APPEARS - If
Fct. 3.6.2 P-SENSOR
is
YES
and
fluid is not
LIQUID.
Fct. 3.7.4 T2
calibration temperature 2
Enter the calibration temperature 2. Unit is assumed
same as selected in
Fct.3.7.1.
LIMITS - 101 to 200 Deg. C.
APPEARS - If
Fct. 3.6.2 P-SENSOR
is
YES
and
fluid is not
LIQUID.
Fct. 3.7.5 P21 V21
calibration point 1 at T2
Enter pressure value (unit assumed to be same as
selected in
Fct. 3.7.2
) and the corresponding milli-
volt value.
LIMITS - 0 to 40 Bar for P21.
0 to 25 mV for V21.
APPEARS - If
Fct. 3.6.2 P-SENSOR
is
YES
and
fluid is not
LIQUID.
Fct. 3.7.6 P22 V22
Calibration point 2 at T2
Enter pressure value (same unit assumed as in
Fct.3.7.2)
and the corresponding milli-volt value.
LIMITS - 5 to 100 Bar for P22
40 to 125 mV for V22
APPEARS - if
Fct. 3.6.2 P-SENSOR
is
YES
and
fluid is not
LIQUID.
Fct. 3.8.0 HEAT PARA.
This submenu groups functions which allow to
view/enter the heat meter power/energy calculation
related data. These functions give the following
information to the instrument
- Units for power
- Units for energy
- Energy totalizer Reset/Set
- Energy totalizer ON/OFF
- Dryness factor for Saturated steam applications.
APPEARS - If
Fct. 1.1.1
MEAS. INST
is
MASS
and
Fct. 3.5.1 FLUID
is
STEAM
or
LIQUID.
Fct. 3.8.1 POWER. UNITS
thermal power units
34