6
OPERATION
96
IFC 300
www.krohne.com
06/2009 - 4000069802 - MA IFC 300 R03 en
C1.1.9
density
Calculation of mass flow with constant density of product;
range: 0.1
…
5 kg/l
C1.1.10
target conduct.
Reference value for on-site calibration; range: 1.000...50000
μ
S/cm
C1.1.11
EF electr. factor
For calculation of the conductivity based on the electrode impedance.
Query: calibrate EF?
break (return with
^
key)
Set value in the following:
standard (with factory setting) /
manual (set desired value) /
automatic (determines EF according to the setting in Fct. C1.1.10)
C1.1.12
num. of electrodes
Selection see measuring sensor nameplate:
2 electrodes (standard) /
3 electrodes (with full pipe electrode) /
4 electrodes (full pipe and grounding electrode)
C1.1.13
field frequency
Setting as on measuring sensor nameplate =
line frequency x value (from the following list):
2; 4/3; 2/3; 1/2; 1/4; 1/6; 1/8; 1/12; 1/18; 1/36; 1/50
C1.1.14
select settling
Select settling (special function)
Select: standard (fixed allocation) /
manual (manual time setting for the settling time for the field current)
C1.1.15
settling time
Only when "manual" selected in Fct. C1.1.14; range: 1.0
…
250 ms
C1.1.16
line frequency
Set line frequency.
automatic (measuring & setting; for DC systems fixed setting 50 Hz)
Select: 50 Hz or 60 Hz (fixed setting)
C1.1.17
act. coil resistance
Display of the current value for calculation of the temperature.
No.
Function
Settings / descriptions
.book Page 96 Wednesday, June 10, 2009 4:20 PM