OPERATION
6
99
IFC 300
www.krohne.com
06/2009 - 4000069802 - MA IFC 300 R03 en
C1.3.17
diagnosis value
Select diagnosis value for testing the various analogue outputs.
Select: off (no diagnosis) / electrode noise (activate Fct. C1.3.13)
flow profile (activate Fct. C1.3.10) / linearity (activate Fct. C1.3.6)
terminal 2 DC (electrode DC voltage) / terminal 3 DC (electrode DC voltage)
C1.4 information
C1.4
information
C1.4.1
liner
Shows material of the liner.
C1.4.2
electr. material
Shows material of the electrodes.
C1.4.3
calibration date
Not available at this time.
C1.4.4
serial no. sensor
Shows serial no. of the measuring sensor.
C1.4.5
V no. sensor
Shows the order number of the measuring sensor.
C1.4.6
sensor electr. info
Shows the production and calibration date of the electronics and the
software version.
C1.5 simulation
C1.5
simulation
C1.5.1
flow speed
Sequence see Fct. B1.1
C1.5.2
volume flow
Sequence see Fct. B1.2
No.
Function
Settings / descriptions
C2 inputs/outputs (I/Os)
C2.1 hardware
C2.1
hardware
Assignment of connection terminals dependent on signal converter version:
active / passive / NAMUR
C2.1.1
terminal A
Select: off (switched off) / current output / frequency output / pulse output /
status output / limit switch / control input / current input
C2.1.2
terminal B
Select: off (switched off) / current output / frequency output / pulse output /
status output / limit switch / control input / current input
C2.1.3
terminal C
Select: off (switched off) / current output / status output / limit switch
C2.1.4
terminal D
Select: off (switched off) / frequency output / pulse output / status output /
limit switch