4
ELECTRICAL CONNECTIONS
32
IFC 300
www.krohne.com
06/2009 - 4000069802 - MA IFC 300 R03 en
1
Strip the conductor to dimension a.
2
Trim the outer shield to dimension b and pull it over the outer sheath.
3
Cut off the inner shield (10), the stranded drain wire (6) and the shields of the insulated con-
ductors. Make sure not to damage the stranded drain wires (1, 20, 30).
4
Slide the insulation tubing over the stranded drain wires (1, 20, 30).
5
Crimp the wire end ferrules onto the conductors and stranded drain wires.
6
Pull the heat-shrinkable tubing over the prepared signal cable.
Figure 4-6: Signal cable B, preparation for field housing
a = 80 mm / 3.15"
b = 10 mm / 0.39"
.book Page 32 Wednesday, June 10, 2009 4:20 PM