19
19
Installation/Connection
About Display
CH00 EXT P1A
[MINOLTA ]
* This shows when the entire display area is lit. (The LCD display section is not shown.)
1 Measurement mode
indications
2 Digital display section
3 Analog display section
4 LCD display section:
1
Measurement mode ................................. Displays the measurement mode in which the measured values are
displayed.
Measurement mode switches from one to another as shown below
each time the MODE key is pressed. (Page 40)
When LED Universal Measuring ø27 Probe(CA-PU32/35) or LED
Universal Measuring ø10 Probe(CA-PSU32/35) is connected,
Analyzer mode
xyLv
→
T
∆
uvLv
→
RBG (R/G, B/G,
∆
G)
→
RBG (
∆
R, B/G, G/R)
→
u'v'Lv
→
XYZ
→
xyLv
When LED Flicker Measuring ø27 Probe(CA-P32/35) or LED Flicker
Measuring ø10 Probe(CA-PS32/35) is connected,
Analyzer mode
xyLv
→
T
∆
uvLv
→
RBG (R/G, B/G,
∆
G)
→
RBG (
∆
R, B/G, G/R)
→
u'v'Lv
→
Flicker**
→
XYZ
→
xyLv
2
Digital display section ............................. Displays the measured values.
●
When xyLv measurement mode is selected
x, y and Lv are displayed.
●
When T
∆
uvLv measurement mode is selected
T,
∆
uv and Lv are displayed.
T (correlated color temperature) is displayed in three significant digits.
●
When an analyzer measurement mode is selected
R, B and G are displayed. R-reference and G-reference are available.
(The same contents are displayed in the digital display area, whether
R-reference or G-reference.)
●
When u'v'Lv measurement mode is selected
u', v' and Lv are displayed.
●
When flicker measurement mode** is selected
Flicker** is displayed. The display range is from 0.0 to 999.9%.
●
When XYZ measurement mode is selected
X, Y and Z are displayed. (X, Y and Z from top to bottom)
x
y
Lv
∆
x
∆
y
∆
Lv
T
Lv
∆
uv
∆
x
∆
y
∆
Lv
R
B
G
R/G
B/G
∆
G
u'
Lv
v'
∆
x
∆
y
∆
Lv
∆
x
∆
y
∆
Lv