DARKEN PROBE
PUSH 0-CAL KEY
ZERO CALIBRATION
CH00 EXT Ad P1A
[ ]
CH01 EXT P3A
[ ]
MENU : SELECT
PUSH SPACE KEY
SELECT : PROBE
P1 35881112 A
SELECT : PROBE
P3 35881113
Message displayed
when the POWER
switch is set to ON
Press the 0-CAL key.
During zero calibration
End of zero calibration
Memory channel
Probe no.
Menu selection screen
PROBE selection screen
Probe no.
Press the
key until the
desired probe
no. appears.
[Operating Procedure]
1
. Set the POWER switch to ON.
2
. Set the 0-CAL ring of the measuring probe
to the 0-CAL position.
Never direct the measuring probe toward a high-inten-
sity illuminant.
When the optional 4-Probe Expansion Board CA-B15 is used
Set the 0-CAL ring of every measuring probe to the 0-
CAL position. Zero calibration will not be performed
correctly if the 0-CAL ring of any of the measuring
probes is not set to the 0-CAL position.
3
. Press the 0-CAL key.
After zero calibration is complete, set the 0-CAL ring
to the MEAS position and start measurement.
4
. Press the MODE key to select analyzer measure-
ment mode (RGB).
5
. Press the MEMORY CH
and
keys to select
the memory channel for which the RGB emission
characteristic for analyzer mode has been set (page
59).
When the optional 4-Probe Expansion Board CA-B15 is used
Select the probe no. for which the RGB emission characteristic for the analyzer mode has been input.
1
Press the key.
The LCD display section will switch to the menu selection screen.
2
Press the key to open the PROBE selection screen.
Each time the
key is pressed, the screen will switch in the
order PROBE
→
SYNC
→
ID Name input
→
RANGE
→
Mea-
surement Speed
→
Number of Digits
→
Calibration Standard
→
RS232C Baud Rate
→
PROBE.
3
Press the key to display the probe no. you want to select.
Each time the
key is pressed, the probe no. switches in the
order [P1] ….
4
Press the key to confirm the selection.
* By default (factory setting), the instrument is set so that [P1] will be selected automati-
cally when the POWER switch is set to ON( | ).
6 .
Place the measuring probe against the display and
take measurement.
Place the probe against the display
Block entry of light