105
Explanation Section
(Note)
•
The
key is inoperable if error message *3 is displayed.
•
The instrument operates as follows if error message *4 is displayed.
1
Clears the display by the MR key.
2
Aborts CAL ON state (i.e. the state that is effective when the
key is pressed).
3
Aborts MENU ON state (i.e. the state that is effective when the
key is pressed).
4
Aborts measuring probe selection.
5
Aborts SYNC mode selection.
6
Aborts ID name setting.
7
Aborts analog display range setting.
8
Aborts measurement speed selection.
9
Aborts selection of the number of display digits.
!0
Aborts calibration standard selection.
!1
Aborts RS-232C baud rate selection.
•
The instrument operates as follows if the error message *5 is displayed.
1
The
key is inoperable during CAL ON state (i.e. the state that is effective when the
key is pressed).
2
The HOLD key is inoperable.
Error Message
Cause: (Description)
Corrective Action
•
Connect the correct probe or expansion
board.
•
The connected probe or expansion board
differs from the one used on the instrument.
(The indication in italics shows the model
name of the probe or expansion board.)
INCORRECT PROBE
PROBE:
INCORRECT
PROBE
PROBE
:
CA100Plus
INCORRECT BOARD
BOARD:
CA100Plus
CA100Plus
INCORRECT
BOARD
BOARD
:
CA100Plus