113
Explanation Section
Item
LED Flicker Measuring Ø10 Probe
Receptor
Detector: Silicon photo cell
Measurement area
Ø10 mm
Acceptance angle
±5°
Measurement distance
30±5 mm
Display range
Luminance
0.0001 to 3000 cd/m
2
Chromaticity
Displayed in 4 or 3-digit value (Can be chosen)
Luminance
Measurement range
0.0150 to 3000 cd/m
2
Accuracy
(for white)*1
Repeatability(2 ) *1
0.0150 to 0.2999 cd/m
2
1% + 0.0030 cd/m
2
0.0150 to 0.2999 cd/m
2
±4%±0.0045
cd/m
2
30.00 to 3000 cd/m
2
±2%±0.0030
cd/m
2
0.3000 to 29.99 cd/m
2
±3%±0.0030
cd/m
2
0.3000 to 2.999 cd/m
2
0.2% + 0.0030 cd/m
2
3.000 to 3000 cd/m
2
0.1% + 0.0030 cd/m
2
Chromaticity
Measurement range
0.1500 to 3000 cd/m
2
Accuracy *1
(temperature:23°±2°,
relative humidity:
(40±10)%))
0.1500 to 14.99 cd/m
2
±0.005 for white
0.1500 to 0.2999 cd/m
2
0.010
0.3000 to 0.5999 cd/m
2
0.004
0.6000 to 1.499 cd/m
2
0.002
1.500 to 3000 cd/m
2
0.001
120 cd/m
2
±0.002 for white
15 cd/m
2
or higher
0.0 ~ 999.9 %
±1 % (Flicker frequency: 30 Hz AC/DC 10% sine wave) ±2 % (Flicker frequency: 60 Hz AC/DC 10% sine wave)
1 % (Flicker frequency: 20 to 65 Hz AC/DC 10% sine wave)
15 cd/m
2
or higher
16 (16) times/sec.
0.5 (0.3) times/sec. *5
60.00 to 3000 cd/m
2
±0.003 for white
15.00 to 59.99 cd/m
2
±0.004 for white
(±0.004 for monochrome)*2
Repeatability(2 ) *1
Measurement range
Display range
Accuracy
Repeatability(2 )
Flicker JEITA
method *3
Measurement range
Accuracy
Repeatability(2 )
Measurement
speed*4
xyL
V
0.0150 to 0.2999 cd/m
2
4(3.5)
times/sec.
6.000 to 3000 cd/m
2
20(17)
times/sec.
0.3000 to 5.999 cd/m
2
5(4.5)
times/sec.
Flicker Contrast
Flicker JEITA *3
Display
Digital
xyL
V
, TΔuvL
V
, RGB analyze, XYZ, u'v'L
V
, Flicker (Contrast method) *3
Analog
ΔxΔyΔL
V
, R/G B/G ΔG, ΔR B/R G/R, Flicker (Contrast method) *3
LCD
16 characters by 2 lines (with backlight)
SYNC mode
NTSC, PAL, EXT, UNIV, INT
Object under measurement
Vertical synchronization frequency: 40 to 200 Hz (Luminance or chromaticity measurement), 40 to 130 Hz (Flicker measurement)
Memory channel
100 channels
Analyzer function
Standard function
Interface
USB; RS-232C (38,400 bps or below)
Multi-point Measurement
Max. 5 points (Use 4-Probe Expansion Board CA-B15)
Operating temperature/humidity range
Temperature: 10 to 28°C; relative humidity 70% or less with no condensation Luminance change: ±2% of reading for white ;
Chromaticity change ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta’s standard LCD *1, 120 cd/m
2
,
with 23°C 40%
Storage temperature/humidity range
0 to 28°C: relative humidity 70% or less with no condensation 28 to 40°C: relative humidity 40% or less with no condensation
Input voltage range
100-240V
AC power cord, PC Software for Color Analyzer CA-SDK, Small Hood CA-HS10
4-Probe Expansion Board CA-B15, LED Flicker Measuring ø10 Probe CA-PS32(2 m) CA-PS35(5 m), Small Hood CA-HS10,
Small Lens Cap CA-HS11, USB cable IF-A27
, 50-60 Hz, 50 VA *6
Size/weight
Standard accessories
Optional accessories
Main body
340(W)×127(H)×216(D) mm / 3.58 kg
Probe
Ø49×236 mm / 550 g
*1 The chromaticity and luminance are measured under Konica Minolta's condition (standard LCD(6500K, 9300K) is used).
*2 The luminance for monochrome is measured when the reading of luminance for white is 120 cd/m
2
*3 Measurement of flicker (JEITA method) is supported by SDK software.
*4 Measuring probe connected to probe connector P1 only, used USB(used RS-232C Baud rate: 38400 bps)
*5 Measured by Konica Minolta's PC (CPU: 600 MHz)
*6 " " represents alternating current available at power outlets.
The specifications and drawings given here are subject to change without prior notice.
Flicker
Contrast
method
±0.5 dB
(Flicker frequency: 30 Hz AC/DC 4%
(-40 dB) sine wave)
±1.0 dB
(Flicker frequency: 30 Hz AC/DC 1.2%
(-50 dB) sine wave)
0.1 dB
(Flicker frequency: 30 Hz AC/DC 4% (-40 dB) sine wave)
0.3 dB
(Flicker frequency: 30 Hz AC/DC 1.2% (-50 dB) sine wave)
Summary of Contents for CA-310
Page 24: ...22...
Page 46: ...44...
Page 86: ...84...
Page 122: ...BGLFPK Printed in Japan 9222 A3CT 11 2010 2016 KONICA MINOLTA INC...