62
Setting/Changing the Target Color
If you have input the RGB emission characteristic for user calibration/analyzer
mode:
It is not necessary to set the target color in the following cases.
1
When you want to set the user-calibrated color as the target color for a memory channel
2
When you want to set the target color (W) which was set when the RGB emission characteristic for the display’s
analyzer mode was set as the target color
The target color set here is the same as those set by
1
and
2
Only when you want to change the currently set target
color, should it be changed it as explained below.
By setting the target color, the difference between the measured value and the target color can be displayed in the
analog display section. The target color can be set for each probe of each memory channel.
The target color must be set in the following cases.
•
When you want to set the target color for memory channel CH00
•
When you want to perform measurement using Konica Minolta’s calibration standard without user calibration
and want to use the analog display function
•
When you want to set a color that differs from the color used for user calibration as the target color to a user-
calibrated memory channel
•
When the optional 4-Probe Expansion Board
CA-B15 is used
When you want to set the target color (W) that has already been set and another color as the target color to a
memory channel for which the RGB emission characteristic for analyzer mode has been input
There are the following two methods of setting/changing the target color. Some memory channels do not allow you
to set the target color.
1. Setting/changing the target ...................... The display’s measured value is set as the target color.
color by measurement
This method can be used for any memory channels.
2. Setting/changing the target color ............. Set the desired values (x, y, Lv) by entering them directly using the
by entering values
instrument’s number-key. This method can be used for memory chan-
nel CH00 only. (This method is not possible if analyzer measure-
ment mode is selected.)
Summary of Contents for CA-310
Page 24: ...22...
Page 46: ...44...
Page 86: ...84...
Page 122: ...BGLFPK Printed in Japan 9222 A3CT 11 2010 2016 KONICA MINOLTA INC...