79
In the alarm menu you can configure flow monitoring. When you have
set CONTROL to FLOW, you can specify 2 additional limit values for
maximum and minimum flow.
If the measured value lies outside this range, an alarm message and a
22-mA error signal (if configured) will be generated.
Configuration
Menu item
Action
Choices
Select function of
CONTROL input
Select using
keys.
Press enter to confirm.
PARSET
(selecting parameter
set A/B via signal at
CONTROL input)
Flow
(for connecting a pulse-
output flow meter)
Adjust to flow meter:
With “Flow” selected,
you must adjust the
device to the flow meter
used.
Enter value using arrow
keys. Press enter to
confirm.
12000 pulses/liter
Display
Flow measurement in measuring mode
Display
Flow measurement (sensor monitor)
5