109
Calibration
Zero Calibration in Air / with Calibration Solution
Display
Action
Remark
Select Calibration.
Press enter to proceed.
Select CAL_ZERO
calibration method.
Press enter to proceed.
Ready for calibration.
Hourglass blinks.
Display (3 sec)
Now the device is
in HOLD mode.
Calibration in air.
Edit digits until the lower
display indicates Zero
Calibration with
solution
Edit digits until the lower
display indicates the
solution value
Press enter to proceed.
The device shows the cell
factor (at 25 °C) and the
zero point.
Sensoface is active.
Use the arrow keys to
select:
• MEAS (exit)
• REPEAT
Press enter to proceed.
Exit:
HOLD is deacti-
vated after a short
time.