61
Configuration
Menu item
Action
Choices
CIP / SIP (ISM only)
The following adjustments are possible for ISM sensors:
Cleaning cycles
on/off
Select ON or OFF using
keys.
Activates/deactivates log-
ging in extended logbook
Press enter to confirm.
ON/OFF
Sterilization cycles
on/off
Select ON or OFF using
keys.
Activates/deactivates log-
ging in extended logbook
Press enter to confirm.
ON/OFF
The cleaning and sterilization cycles are logged to measure the load
on the sensor.
Suitable for biochemical applications (process temp approx. 0 ... 50 °C,
CIP temp > 55 °C, SIP temp > 115 °C).
Note:
A CIP or SIP cycle is only entered into the logbook 2 hours after the
start to ensure that the cycle is complete.
5