23
Preparing the shield connection
Pre-assembled special cable for SE 655 / SE 656 sensors
Cable Preparation SE 655 / SE 656
The pre-assembled special cable:
BU
RD
RD
BU
GN
WH
YE
GN/YE
Coax RD
Coax WH
• Insert the special cable through
the cable entry into the terminal
compartment.
• Remove the already separated part
of the cable insulation (1).
• Turn the shielding mesh (2) over the
cable insulation (3).
• Then shift the crimp ring (4) over
the shielding mesh and tighten it
using a pince (5)
3
4
5
receive hi
receive lo
send lo
send hi
RTD (temp)
RTD (temp)
sense (temp)
shield