120
Service
Display
Menu item
Displaying the currently measured values (sensor
monitor) with HOLD mode activated:
Select MONITOR using
, press enter to confirm.
Select variable in the bottom text line using
.
The selected parameter is shown in the main display.
As the device is in HOLD mode, you can perform
validations using simulators without influencing the
signal outputs.
Press meas to return to the service menu.
Return to measurement: Press meas once more.
Specifying the current at outputs 1 and 2:
Select OUT1 or OUT2 using the
keys,
press enter to confirm.
Enter a valid current value for the respective output
using
keys.
Press enter to confirm.
For checking purposes, the actual output current is
shown in the bottom right corner of the display.
Exit by pressing enter or meas.
Relay test (manual test of contacts):
Select RELAIS using
, press enter to confirm.
Now the status of the 4 relays is “frozen“. The 4 digits
of the main display represent the respective states
(from left to right: REL1, REL2, ALARM, WASH).
Select one of the 4 relays using the
keys, close (1)
or open (0) using the
keys.
The selected position blinks. The lower display line
shows the relay function, e.g. “WASH” or “ALARM”.
End by pressing enter. The relays will be re-set
corresponding to the measured value.
Press meas to return to measurement.