37
Maintenance and Care
KNAUER
10D User Manual V7670, Version 2.3
Removing the Fiber Optics
Prerequisites
The device has been switched off.
Tools
Cap fittings for the fiber optics
Note
Do not touch the ends of the fiber optics with your fingers, as
this could falsify the measurement
Procedure
Process
Figure
1. Manually, unscrew the
fittings
1
of the fiber
optic from the flow cell.
Fig. 27
Fitting of the fiber
optics on the flow cell
2. Seal the fiber optic con-
nectors with caps
2
.
Fig. 28
Cap fitting on the fiber
optics
3. Manually unscrew the
fiber optic
3
from the
detector.
Fig. 29
Fitting of the fiber
optics on the detector
1