36
Maintenance and Care
KNAUER
10D User Manual V7670, Version 2.3
Replacing the Flow Cell
Prerequisite
Detector has been switched off.
Power plug has been pulled.
Capillaries are disconnected.
Replacing the Fiber Optic
UV light will cause the optical fiber to become blind with time
(solarization), making them no longer suitable for use.
Observe the following regarding the use of UV fiber optics:
Do not touch the ends of the fiber optics with your fingers, as
this could falsify the measurement.
Handle the fiber optics with care, avoid impacts or hard
actions.
Move the fiber optics carefully without using pressure or
bending it.
Procedure
Steps
Figure
1. Unscrew the 2 knurled-
head screws
1
.
2. Pull out the slide
2
.
3. Remove the flow cell.
Fig. 25
Removing the slide
4. Insert the flow cell
4
into the slide
3
.
5. Push the slide into the
detector.
6. Tighten the knurled-
head screws.
Fig. 26
Assembling the flow cell
Result
Flow cell is assembled. The next step is connecting the capil-
laries.
1
2
3
4