29
Operation
KNAUER
10D User Manual V7670, Version 2.3
Entering a Scaling Value
It is recommended to set 500 nm at an applied voltage of 5 V.
Prerequisite
Device has been connected to GROUND.
ANALOG port has been selected.
Zero point has been set.
Control with Chromatography Soft-
ware
The detector can be controlled with e. g. OpenLAB EZChrom
edition version A.04.05 or higher, ChromGate version 3.3.2 or
higher and ClarityChrom version 3.0.7 or higher. You will find a
detailed description on the chromatography soft-ware in the
software manual.
There are two advantages to the use of software:
Continuous operation for triggered measurement is possible.
System status can be checked in the software.
To be able to control the device using chromatography soft-
ware, the computer must be connected to the device either with
a RS-232 cable or a LAN cable.
Process
Figure
Procedure
1. Apply voltage.
2. Keep left arrow key
pressed. Press right
arrow key until the cor-
rect display appears.
3. Let go of left arrow key.
4. Using the arrow keys,
set a value. Do not
press Autozero.
Fig. 20
Display Analog-In Scal-
ing
Result
When setting is finished, the status display appears.
0nm
A-IN