Initial Startup
KNAUER
10D User Manual V7670, Version 2.3
Initial Startup
Note
Before the detector is ready to use, a flow cell must be installed.
Before installing the flow cell, the optical path length can be
changed. The optimal path length depends on type and quan-
tity of the sample.
Inserting the Flow Cell
The test cell has no connectors for capillaries and must therefore
be replaced by a flow cell before first usage.
Note
The test cell is used during operation qualifications and must be
stored.
Prerequisite
Detector has been switched off.
Power plug is pulled out.
Process
Figure
Procedure
1. Unscrew the 2 knurled-
head screws
1
.
2. Pull out the slide
2
.
3. Remove the test cell.
Fig. 5
Removing the slide
4. Insert the flow cell
4
into the slide
3
.
5. Push the slide into the
detector.
6. Insert the knurled-head
screws and screw tight.
Fig. 6
Assembling the flow cell
Result
Flow cell is assembled. The next step is connecting the capil-
laries.