Unpacking and Setup
KNAUER
10D User Manual V7670, Version 2.3
Next steps
Finish the installation and perform the initial startup.
Analog Port
Using the analog port (ANALOG-IN), you can control the wave-
length by changing the applied voltage. Before connecting the
analog port, you have to choose analog in the setup menu. The
control voltage must be applied against GROUND. (see See
“Using the Analog Port” on page 27.)
Example
To use the analog port for controlling the detector, you have to
set a zero point and enter a scaling value.
Zero point at 0 V = 000 nm
Scaling: 100 nm per Volt
If 5-V voltage is applied, the wavelength is 500 nm.
Ground
Note
The detector is not grounded if it is connected with a 24-V
power supply to a 220-V outlet.
Voltages, which may influence the measuring result, can be dis-
charged if the device is detected to a grounded KNAUER device,
e. g., a thermostat.
Tool
Depressor tool
Procedure
Process
Figure
1. Place the spring strip
3
on a suitable surface.
2. Push the depressor tool
1
into the opening on the
spring strip.
3. Continue pushing the
depressor tool down and
lead the cable
2
into the
front end of the spring
strip.
4. Remove the depressor
tool.
5. Check whether the cables
are tightly attached.
6. Plug the spring strip onto
the pin header.
Fig. 4
Spring strip
1
2
3