Functionality Tests
KNAUER
10D User Manual V7670, Version 2.3
Functionality Tests
Installation Qualifica-
tion (IQ)
Certification on the functionality of the device. During installa-
tion of the device, an installation report (
IQ
document) is cre-
ated upon request in coordination with the Technical Support of
KNAUER.
This installation report needs to be completed in full and signed
by both parties. It serves as proof of the properly executed
installation and functionality of the device.
Operation Qualification
(OQ)
Extensive test of the device’s functionality. A successfully exe-
cuted
OQ
ensures that the device functions properly.
Test Intervals
Run the extensive functionality test at the following time inter-
vals:
Execution
The test can carried out either by KNAUER's Technical Support
or by a technical service authorized by KNAUER.
The
OQ
documentation required for executing the
OQ
is with
costs (once) and can be ordered separately from KNAUER.
Average use
Device test
1 to 5 days/week:
Every 6
months
More than 5 days/week or 24 hours/day:
Every 3
months
Operation with buffer solutions or other salt
solutions:
Every 3
months