Technical data
4000 MiD® Instructions, V6695
11. Technical data
Requirement
Description
Indoor use
The 4000 MiD can only be used in an indoor environ-
ment
Bench space
Required:
56cm x 35.5cm x 25.5cm (L x W x H)
MiD weight:
32 kg
Power
requirements
Line voltage:
100–240V AC ±10 %
Line Frequency: 50/60 Hz
Power:
300W (peak)
Overvoltage
Category II:
For transient overvoltages typically
present on the mains supply
For temporary overvoltages on the
mains supply
Altitude
Operation:
Up to 2000 m
Temperature
Operation:
15–30
°
C
Storage:
-10–50
°
C
Relative
humidity
Operation:
40–80 %
Pollution
Degree 2:
Normally only non-conductive
pollution occurs. Occasionally,
however, a temporary conductivity
caused by condensation must be
expected
Storage:
10–90 %, non-condensing
Nitrogen gas
Pressure:
2–6 bar (29–87 psi)
Purity:
≥ 99.5 %
Tubing:
Site must use either stainless steel
or PTFE tubing
Fitting:
Stainless Steel
∅
6 mm push lock
fitting
Pump exhaust
Tubing:
∅
8 mm push lock fitting
12. Reorder
Part
Article no.
Spraychip 200
A66901
All metal vac-chip
A66902
Split flow interface (SFI)
A66913
Re-built kit for split flow interface
A66914
Calibration kit for split flow interface
A66916