21
Product information
4000 MiD® Instructions, V6695
Fig. 8
MiD without the vac-chip assembly fitted
2.
Locate the vac-chip flange on the SFD using the dowels as shown in
Fig. 9. Gently push the vac-chip flange into place, do not force the
flange into position.
Note:
There is a single orientation where the holes and
threaded sections will align.
Fig. 9
The vac-chip assembly being installed on the MiD
3.
If excessive resistance is felt when fitting the assembly to the SFD
stop immediately, remove the assembly, and retry step 2.
4.
Once positioned on the SFD fit the four M4 screws provided.
5.
Use the 4 mm “T” Allen key and tighten the four M4 screws in an
alternating criss-cross pattern until the screws are tight.
6.
If excessive force is required stop immediately and repeat steps 2-5.
7.
Once the vac-chip assembly is fitted the MiD should look as shown in
Dowels
vac-chip assembly