90
PWX
Calibration
The PWX series is calibrated before shipment. To maintain long-term performance, we rec-
ommend periodic calibration.
For calibration, contact your Kikusui agent or distributor.
If you are going to calibrate the PWX series yourself, follow the procedures below. All of the
calibration items of the PWX series are described.
Calibration overview
The following eight calibration items are available.
Be sure to calibrate both the offset and full scale values.
•
Output voltage offset (10 %)
• Output voltage full scale (100 %)
•
Voltmeter offset (0 %)
• Voltmeter full scale (100 %)
•
Output current offset (10 %)
• Output current full scale (100 %)
•
Ammeter offset (0 %)
• Ammeter full scale (100 %)
Required devices
•
DC voltmeter (DVM) with a measurement accuracy of 0.02 % or more
•
Shunt resistor
Environment
Perform calibration in the following environment.
•
Temperature: 23 °C ± 5 °C
•
Humidity: 80 %rh or less
To minimize the calibration error due to initial drift, warm up the PWX series for at least 30
minutes before calibration. In addition, warm up the DVM and shunt resistor for as long as is
necessary.
Model
Recommended shunt resistor
Rating
Tolerance
PWX750LF
100 A / 50 mV (0.50 m
Ω
)
< ±0.1 %
PWX750MLF
30 A / 50 mV (1.67 m
Ω
)
PWX750MHF
10 A / 50 mV (5 m
Ω
)
PWX750HF
5 A / 50 mV (10 m
Ω
)
PWX1500L
150 A / 50 mV (0.50 m
Ω
)
PWX1500ML
100 A / 50 mV (0.50 m
Ω
)
PWX1500MH
20 A / 50 mV (2.5 m
Ω
)
PWX1500H
10 A / 50 mV (5 m
Ω
)
Summary of Contents for PWX1500H
Page 10: ...10 PWX This page is intentionally blank...
Page 26: ...26 PWX This page is intentionally blank...
Page 88: ...88 PWX This page is intentionally blank...
Page 89: ...Maintenance This chapter explains how to perform cali bration DELETE...
Page 92: ...92 PWX This page is intentionally blank...
Page 120: ...120 PWX This page is intentionally blank...
Page 121: ...Appendix A Options B Troubleshooting...