76
PWX
Pin no.
Signal name
Description
Circuit
1
STATUS COM
This is the status output common.
This is the photocoupler emitter output.
2
CV STATUS
This is set to low level when the PWX series is in constant
voltage mode.
This is the photocoupler open collector output.
3
CC STATUS
This is set to low level when the PWX series is in constant
current mode.
This is the photocoupler open collector output.
14
ALM STATUS
This is set to low level when a protection function (OVP/
OVP2/ OCP/ OHP/ OHP2/ FAN/ SEN/ AC-FAIL/ SD) is
activated.
This is the photocoupler open collector output.
15
PWR ON
STATUS
This is set to low level when the PWX series is turned on
(PWR ON STATUS).
This is the photocoupler open collector output.
16
OUT ON STATUS This is set to low level when the output is turned on.
This is the photocoupler open collector output.
1
2
3
14
15
16