Calibration
5
Maintena
nce
Calibration procedure
Use the “KIKUSUI PWX calibration” application software included on the CD-ROM to cali-
brate the PWX series. The system requirements for the “KIKUSUI PWX calibration” is below.
OS
•
Windows7 (x86/ x64)
•
Windows Vista (x86/ x64)
•
Windows XP—SP2 (x86), .NET Framework 2.0 or later
VISA library
One of the following VISA libraries is required. Do not install multiple VISA libraries on the
same PC. Doing so may cause errors.
•
NI-VISA by National Instruments Corporation (Ver. 5.0.3 or later)
•
Agilent VISA by Agilent Technologies (Agilent IO Libraries Suite 16.0 or later)
•
KI-VISA Ver. 5.0.4 or later
For details, see the PWX Calibration—Manual that is installed together with the application
software for calibrating the PWX series.
Be sure to go through all the calibration items. If you cancel the calibration item in progress to
initiate a different item or if you turn the POWER switch off, the calibration will be invalid.
Connection
•
Voltage calibration
•
Current calibration
DVM
HI
LO
Output terminal
Chassis terminal
PWX
Shunt
DVM
HI
LO
Output terminal
Chassis terminal
PWX