Troubleshooting
This section introduces troubleshooting measures. Typical symptoms are listed. Check
whether any of the items listed below apply to your case. In some cases, the problem can be
solved quite easily.
If none of the items apply to your case, we recommend that you initialize the KPM1000 to its factory
default settings. If the corrective action does not correct the problem, Contact your Kikusui agent or
distributor..
The power does not turn on
Unable to carry out panel operations
Remote interface
See
Symptom
Item to check
Corrective action
The KPM1000 does not operate
when the POWER switch is turned
on.
Is the power cord
disconnected?
Connect the power cord correctly.
See
Symptom
Item to check
Corrective action
Unable to operate the panel keys.
Is the KEY LOCK LED status
indicator lit?
Press the KEY LOCK key to release the key
lock.
Is the RMT LED lit?
The KPM1000 is being operated through the
RS232C, GPIB, or USB interface. To control
the KPM1000 from the panel, press the
LOCAL key to switch to local mode.
Unable to change range,
configuration, or interface settings.
Is the HOLD LED lit?
Values are being held. Press the HOLD key to
stop holding values.
Is the KEY LOCK LED status
indicator lit?
Press the KEY LOCK key to release the key
lock.
Is the START STOP LED lit?
Integration is being performed. When the
KPM1000 is in this state, you cannot change
settings; you can only view them.
Is the START STOP LED
blinking?
Integration has been paused. When the
KPM1000 is in this state, you cannot change
settings; you can only view them.
Even though the auto range
feature is turned on, pressing the
START STOP key turns the auto
range feature off.
Is the START STOP LED lit?
When integration is being performed, you
cannot use the auto range feature. It will be
turned off automatically.
See
Symptom
Item to check
Corrective action
The remote interface factory
default setting is different between
multiple KPM1000s.
Is a communication interface
factory option installed in the
KPM1000?
The interface that is selected in the factory
default settings is the optional interface.
Even though I changed the remote interface setting, I cannot
perform communications using the specified interface.
Change the setting on the configuration menu,
and then restart the KPM1000.