KPM1000
Basic Operations
21
Displayed contents
Displays B and D
Display item
Symbol
display
Unit display
(center column)
Unit display
(right column)
Description
Voltage
V
-
Values that are 0.5 % of the voltage
range or less are displayed as
0.0000 (they are muted).
Current
A
-
Values that are 0.5 % of the current
range or less are displayed as
0.0000 (they are muted).
Active power
W
-
This is the factory default setting for
display D.
Apparent power
-
-
VA
If both the voltage and current are
muted, this displays 0.0000.
This is the factory default setting for
display B.
Reactive power
-
-
var
If both the measured voltage and
measured current are muted, this
displays 0.0000.
Integrated current
(sum)
-
A
h
This displays the sum of the positive
and negative integrated currents.
If there is no integrated value, “-----”
is displayed.
Integrated power
(sum)
-
W
h
This displays the sum of the positive
and negative integrated powers.
If there is no integrated value, “-----”
is displayed.
Integrated power
(positive)
P
W
h
This displays the positive integrated
power.
If there is no integrated value, “-----”
is displayed.
Integrated power
(negative)
n
W
h
This displays the negative integrated
power.
If there is no integrated value, “-----”
is displayed.
Integration time
-
-
h
Up to 99 hours, 59 minutes, and 59
seconds, this displays the hours,
minutes, and seconds (in the
following format: 99h59.99).
From that point to 9999 hours and
59 minutes, this displays the hours
and minutes (in the following format:
9999h59).
When integration has been cleared,
"-----" is displayed.
Voltage crest factor
C
V
1
1
This displays the voltage.
-
When the voltage is muted, “-----” is
displayed.
Current crest factor
C
A
2
2
This displays the current.
-
When the current is muted, “-----” is
displayed.
Phase angle
d
-
-
When the voltage is muted, "-----" is
displayed.