KPM1000
Specifications
41
Power measurement feature
Voltage measurement range and the display digits
The active power is displayed in units of W. For apparent power and reactive power, read the
unit “W” in the following table as “VA” and “var,” respectively.
Item
Specification
Measurement range
(combination of the
voltage and current
ranges)
Resolution
750 mW
0.01 mW
1.5 W
0.1 mW
3 W
6 W
7.5 W
15 W
1 mW
30 W
60 W
75 W
150 W
0.01 W
300 W
600 W
750 W
1.5 kW
0.1 W
3 kW
6 kW
Accuracy
1
,
2
1
Under the following conditions: within the effective input range, within six months after the last calibration,
at a temperature of 23 °C ± 5 °C, sine wave, power factor of 1, and common mode voltage of 0 V.
If the waveform is asymmetrical (waveforms such as half-wave rectified waveforms and full-wave rectified
waveforms), errors will occur.
2
When you are using the 5 mA range and the input is less than 10 % of the range, add (0.1 % of range).
Effective input
range
Full range
1 % to 144 % of the range
45 Hz
≤
f
≤
66 Hz,
LPF: On or off
±(0.1 % of r 0.1 % of range)
66 Hz < f
≤
400
Hz, LPF: Off
±(0.1 % of r 0.2 % of range)
400 Hz < f
≤
1
kHz, LPF: Off
±(0.1 % of r 0.3 % of range)
1 kHz < f
≤
3 kHz
LPF: Off
±(3 % of range)
Influence of the power
factor
Power factor 0
45 Hz to 66 Hz
±0.4 % of VA
0 < power factor <
1
Add (tan
θ
× 0.4 )% of reading
One-year accuracy (accuracy up to 12 months after calibration)
1.5 times the reading errors for the accuracy
at 6 months
I range
V range
5.0000 mA
10.000 mA
20.000 mA
50.000 mA
100.00 mA
200.00 mA
150.00 V
750.00 mW
1.5000 W
3.0000 W
7.5000 W
15.000 W
30.000 W
300.00 V
1.5000 W
3.0000 W
6.0000 W
15.000 W
30.000 W
60.000 W
I range
V range
500.00 mA
1.0000 A
2.0000 A
5.0000 A
10.000 A
20.000 A
150.00 V
75.000 W
150.00 W
300.00 W
750.00 W
1.5000 kW
3.0000 kW
300.00 V
150.00 W
300.00 W
600.00 W
1.5000 kW
3.0000 kW
6.0000 kW
Summary of Contents for IB022853
Page 5: ...KPM1000 5...
Page 51: ......