Troubleshooting
I cannot measure frequencies.
Is the synchronization source
setting set to “oFF?”
Check the SynC setting on the configuration
menu.
Is the synchronization source
setting set to “Curr” even
though no current is flowing?
Are you trying to measure a
frequency of 500 Hz or more
with the frequency filter turned
on?
Check the F.FiLt setting on the configuration
menu. Change the setting to “oFF” if
necessary.
Is the frequency that you are
applying outside of the
measurement range (10 Hz to
10 kHz)?
Use an external measuring instrument to check
the frequency.
-
Integration starts, but it stops
after a certain amount of time
has elapsed.
Is the integration time correct?
Check the intEG setting on the configuration
menu.
During integration, the Wh and
Ah indicators are blinking.
Is an over-range occurring
during integration?
Set the range to a higher range, and then start
integration again.
Symptom
Item to check
Corrective action
See