38
Specifications
KPM1000
Integrated current
measurement:
Ah (sum)
Status indicators
The START/STOP LED is lit during integration. This LED blinks
when integration is paused and turns off when integration is
reset.
Display range
±0.00000 m to ±999999 M
Mute: Not used.
If the integrated value overflows, “OVEr.F” is displayed.
When the integrated value does not exist, “-----” is displayed.
Error warning
If the integrated values contain values that would cause an
over-range indicator to appear, the unit blinks.
Integrated power
measurement:
Wh (sum),
[P]Wh (positive),
[n]Wh (negative)
Status indicators
The START/STOP LED is lit during integration. This LED blinks
when integration is paused and turns off when integration is
reset.
Display range
±0.00000 m to ±999999 M
Mute: Not used.
If the integrated value overflows, “OVEr.F” is displayed.
When the integrated value does not exist (after INTEG
RESET), “-----” is displayed.
Error warning
If the integrated values contain values that would cause an
over-range indicator to appear, the unit blinks.
Voltage crest factor: [C]V
Display range
0.0000 to 99.999
If the voltage is muted, “-----” is displayed.
Over-range indicators
If a voltage peak over-range or RMS value over-range occurs:
oVEr.
Current crest factor: [C]A
Display range
0.0000 to 99.999
If the current is muted, “-----” is displayed.
Over-range indicators
If a current peak over-range or RMS value over-range occurs:
oVEr.
Frequency measurement:
Hz
Display range
10.000 Hz to 10.000 kHz
You can select to display either the voltage frequency or the
current frequency.
When neither frequency is selected, “-----” is displayed.
Over-range indicators
If a voltage over-range occurs or if the voltage is muted, “-----”
is displayed.
If the measured frequency is greater than the upper limit of the
measurement range, “oVEr” is displayed. If the measured
frequency is less than the lower limit of the measurement
range, “undEr” is displayed.
Measurement interval
Display update interval.
Voltage peak
measurement: Vpk
Display range
Three times the voltage range.
The peak displays the positive or negative value, whichever is
larger.
(The scaling feature can be used to display values in the range
of 0.0000 m to ±99999 M.)
Over-range indicators
If a value exceeds three times the range, the over-range
indicators turn on.
If only a peak over-range occurs: P.oVEr.
If a peak and RMS over-range occur: oVEr.
If only an RMS over-range occurs: Normal display.
Current peak
measurement: Apk
Display range
Six times the current range.
The peak displays the positive or negative value, whichever is
larger.
(The scaling feature can be used to display values in the range
of 0.0000 m to ±99999 M.)
Over-range indicators
If a value exceeds six times the range, the over-range
indicators turn on.
If only a peak over-range occurs: P.oVEr.
If a peak and RMS over-range occur: oVEr.
If only an RMS over-range occurs: Normal display.
Integration time: h
Display range
Up to 99 hours, 59 minutes, and 59 seconds, the hours,
minutes, and seconds are all displayed (99h59.99). From this
point to 9999 hours and 59 minutes, only the hours and
minutes are displayed (9999h59).
When the KPM1000 is not in integration mode, “-----” is
displayed.
Over-range indicators
If the time exceeds 9999 hours and 59 minutes (9999h59),
“oVEr” is displayed.