KPM1000
Installation and Preparation
13
Wiring the measurement circuit
Connecting the voltage input terminal to the load
You can connect the current input terminal to the L side or the N side.
Connecting the current input terminal to the L side
Connecting the current input terminal to the N side
Wire
number
Start point
End point
1
Output terminal G of the AC power supply
Terminal G (ground) of the load
2
Output terminal N of the AC power supply
Terminal N of the load
3
± terminal of the KPM1000 current input
Terminal L of the load
4
Terminal I of the KPM1000 current input
Output terminal L of the AC power supply
5
± terminal of the KPM1000 voltage input
Terminal N of the load
6
Terminal U of the KPM1000 voltage input
Terminal L of the load
±
±
I
U
±
±
I
U
A
V
L
N
G
L
N
L
N
L
N
G
AC power
supply
Load
Load
AC power
supply
KPM1000
Current input terminal
Voltage input terminal
3
1
1
4
2
5
6
6
to
Wire numbers
Wire
number
Start point
End point
1
Output terminal G of the AC power supply
Terminal G (ground) of the load
2
Terminal I of the KPM1000 current input
Terminal N of the load
3
± terminal of the KPM1000 current input
Output terminal N of the AC power supply
4
Output terminal L of the AC power supply
Terminal L of the load
5
± terminal of the KPM1000 voltage input
Terminal I of the KPM1000 current input
6
Terminal U of the KPM1000 voltage input
Terminal L of the load
±
±
I
U
A
V
±
±
I
U
L
N
L
N
Load
AC power
supply
Current input terminal
Voltage input terminal
L
N
G
L
N
G
AC power supply
Load
KPM1000
3
1
1
4
2
5
6
6
to
Wire numbers
Summary of Contents for IB022853
Page 5: ...KPM1000 5...