36
Specifications
KPM1000
Status indicators
OVER V
LED color: Red.
Lights when the peak value of the voltage input waveform
exceeds 3 times the range or when the RMS value exceeds 1.3
times the range.
OVER I
LED color: Red.
Lights when the peak value of the current input waveform
exceeds 6 times the range or when the RMS value exceeds 1.3
times the range.
AVE
LED color: Green.
Lights when the AvE setting on the CONFIG menu is set to a
value between 2 and 64 (when averaging is turned on); turns
off when AvE is set to “off” (averaging is turned off).
LPF
LED color: Green.
Lights when the LPF setting on the CONFIG menu is set to
“on;” turns off when LPF is set to “off.”
CT
LED color: Green.
Lights when the Ct setting on the CONFIG menu is set to “on.”
PT
LED color: Green.
Lights when the Pt setting on the CONFIG menu is set to “on.”
KEY LOCK
LED color: Green.
Lights when the keys are locked.
V RANGE AUTO
LED color: Green.
Lights when the auto voltage range is turned on.
I RANGE AUTO
LED color: Green.
Lights when the auto current range is turned on.
START/STOP
LED color: Green.
Lights when integration starts; blinks when integration has
been paused; turns off when integration is reset.
RMT
LED color: Green.
Lights when the KPM1000 is in remote mode.
HOLD
LED color: Green.
Lights when measured values are being held.
RMS voltage
measurement: V
Display range
0.5 % to 130 % of the range.
(The scaling feature can be used to display values in the range
of 0.0000 m to 99999 M.)
Mute: Values that are 0.5 % of the range or less are displayed
as 0.0000.
Over-range indicators
If a value exceeds 130 % of the range, the over-range
indicators turn on. If an RMS over-range occurs, the buzzer
also sounds.
Over-range indicators:
If only a peak over-range occurs: P.oVEr.
If a peak and RMS over-range occur: oVEr.
If only an RMS over-range occurs: oVEr.
RMS current
measurement: A
Display range
0.5 % to 130 % of the range.
(The scaling feature can be used to display values in the range
of 0.0000 m to 99999 M.)
Mute: Values that are 0.5 % of the range or less are displayed
as 0.0000.
Over-range indicators
If a value exceeds 130 % of the range, the over-range
indicators turn on. If an RMS over-range occurs, the buzzer
also sounds.
Over-range indicators:
If only a peak over-range occurs: P.oVEr.
If a peak and RMS over-range occur: oVEr.
If only an RMS over-range occurs: oVEr.
Item
Specification