ELS-304
5-8
Operations of Load Modes
5.7.1
CV+Climit Mode
In this CV +Climit mode ,by using the curr. limit function in the
configuration setting, you can limit the load current in CV mode.
This is adapted for simulating the charge characteristics of
batteries. You can set the current in about 110% of the value you
set in curr. limit.
Constant current operation with current limit up to setting volt. can be executed.
As an example, in case of batteries, load volt value is the value you setup in CV mode, and at the
same time load current value is determined by the battery volt. and the internal resistance of the
battery.
To set curr. limit, refer to
→
”
5.4
Curr. Limit Setting
”.
・
Confirmed values are retained if you change the load mode.
・
If you input data exceeding the upper limit thereof , the upper limit value is inputted.
Supplement
I
V
Load
Cuurent
Input Volt.
Setup Voltag
Cuurent-Limit