Keysight N5227B Service Guide 3-51
Tests and Adjustments
Adjustments
Figure 3-19
Equipment Setup 3 for the Receiver Adjustment
EE Default Adjustment
This sets the EEPROM data to their default values.
Procedure
1.
Press UTILITY
, then
, then
, then
click EE Default Adjustment.
2.
Follow the instructions and prompts as they are displayed.
IF Response Adjustment (Available Only with Option S93093/4A/B
and or S93900/1/2/4/5/7A/B Installed)
The IF Response adjustment is required when upgrading to Option 090, 093
and or 094. In addition, this adjustment must be performed each time a new
configuration is set up or if cables are changed:
Procedure
1.
Press UTILITY
, then
, then
, then
click IF Response Adjustment.
2.
Follow the instructions and prompts as they are displayed.
System
Service
Adjustment Routines
In A models these options were 090 and 093 or 094.
System
Service
Adjustment Routines