3-46 Keysight N5227B Service Guide
Tests and Adjustments
Adjustments
3-
Step 4.
When the User Account Control opens, press
Yes
.
Step 5.
After pressing all of the cross hair calibration points a Digitizer Calibration
Tool window opens, press
Yes
.
Step 6.
In the Tablet PC Settings window, press
OK
.
Verification Procedure
Step 1.
Verify your instrument is powered up.
Step 2.
Press
System
>
Service
>
Display...
>
Display Setup
>
Calibrate
Touchscreen...
Step 3.
In the Table PC Setting window that opens, press
Calibrate...
Step 4.
Follow the instrument’s display prompts.
Equipment Used for the 10 MHz Frequency Reference Adjustment
Procedure
1.
Connect the equipment as shown in
. Connect a GPIB cable
between the network analyzer and the frequency counter.
Figure 3-15
Equipment Setup for the 10 MHz Frequency Reference Adjustment
If your display fails the verification process, you can reset your display
using the Table PC Setting window. Refer to
.
Equipment Type
Model or Part Number
Alternate Model or Part Number
Cable, BNC, 50
Ω
, 24 inch
8120-1839
Any
Frequency counter
53151A, Option 001
Any that will measure a signal at 10 MHz.
This adjustment typically adjusts to within
±
0.01 ppm.