4-34 Keysight N5227B Service Guide
Troubleshooting
Measurement System Troubleshooting
4-
1. Press UTILITY
, then
, then
. In
the Address box, click the down arrow and then click My Computer.
In an open area of the My Computer window, click the right mouse
button and then click
Properties
in the resulting pop-up menu.
Click the
Hardware
tab, click
Device Manager
, and then expand
Network Adapters
in the resulting list. The following entry should be
listed:
Keysight Technologies DSP Driver #2
and should be enabled.
2. If the entry is not present or if the icon to the left of the name is a
yellow box containing an exclamation mark (!), navigate the following
directories and verify the presence of the following file:
C:\WINNT\system32\drivers\spampnp.sys
.
3. If you have verified that the DSP driver is present and enabled, but all
traces are still missing in all bands, go to
.
—
If the traces exhibit power drops in some frequency bands
, the problem is
“Source Group Tests” on page 4-35
and perform
the tests that correspond to the problems seen.
Single Trace (A, B, C, D, R1, R2, R3, or R4) Only
If the trace is missing in all bands or has notches or roll-off, go to
the Signal Separation Group” on page 4-41
A, B, R1, and R2 Traces Only
The problem is in the source 1 group, go to
“Checking the Source Group” on
C, D, R3, and R4 Traces Only
The problem is in the source 2 group, go to
“Checking the Source Group” on
System
Configure
Control Panel...