Keysight E8357D/67D & E8663D PSG User’s Guide 115
Basic Digital Operation
Using Waveform Markers
Accessing Marker Utilities
Use the following procedure to display the marker parameters. This procedure
uses the Dual ARB player, but you can access the marker utilities through the
ARB Setup
softkey in all ARB formats.
1.
Select the ARB waveform player:
press
Mode
>
Dual ARB
2.
Press
ARB Setup
>
Marker Utilities
.
Sample range begins on first point of signal
Negative range set between signal and
off time
Example of Incorrect Use
Waveform: 1022 points
Marker range: 110–1022
Marker polarity: Negative
This figure shows that a negative polarity marker goes low
during the marker
on
points; the marker signal goes high
during the
off
points. The ALC samples the waveform during
the
off
marker points.
Marker On
Marker On
Sampling both on and off time sets the modulator circuitry
incorrectly for higher signal levels. Note the increased
amplitude at the beginning of the pulse.
Marker
Off
Marker On
Marker
Marker
Off
On