58 Keysight E8357D/67D & E8663D PSG User’s Guide
Basic Operation
Configuring the RF Output
9.
Highlight the frequency item for point 8, then press
Insert Item
.
Pressing
Insert Item
shifts frequency values down one row, beginning at
point 8. Note that the original frequency values for both points 8 and 9
shift down one row, creating an entry for point 10 that contains only a
frequency value (the power and dwell time items do not shift down).
The frequency for point 8 is still active.
10.
Press
590
>
MHz
.
11.
Press
Insert Item
>
–2.5
>
dBm
.
This inserts a new power value at point 8 and shifts down the original
power values for points 8 and 9 by one row.
12.
Highlight the dwell time for point 9, then press
Insert Item
.
A duplicate of the highlighted dwell time is inserted for point 9, shifting
the existing value down to complete the entry for point 10.
To Configure a Single List Sweep
1.
Press
Return
>
Sweep
>
Freq & Ampl
This turns the sweep on again. No errors should occur if all parameters for
every point have been defined in the previous editing process.
2.
Press
Single Sweep
.
The signal generator will single sweep the points in your list. The
SWEEP
annunciator activates during the sweep.
3.
Press
More (1 of 2)
>
Sweep Trigger
>
Trigger Key
.
This sets the sweep trigger to occur when you press the
Trigger
hardkey.
4.
Press
More (2 of 2)
>
Single Sweep
.
This arms the sweep. The
ARMED
annunciator is activated.
5.
Press the
Trigger
hardkey.
The signal generator will single sweep the points in your list and the
SWEEP
annunciator will be activated during the sweep.
To Configure a Continuous List Sweep
Press
Sweep Repeat Single Cont
.
This toggles the sweep from single to continuous. A continuous repetition of
the frequencies and amplitudes configured in the list sweep are now available
at the RF OUTPUT connector. The
SWEEP
annunciator appears on the display,
indicating that the signal generator is sweeping and progression of the sweep
is shown by a progress bar.