Keysight E8357D/67D & E8663D PSG User’s Guide 311
Peripheral Devices
Millimeter-Wave Source Modules
1.
Turn on the signal generator’s line power.
NOTE
A
utomatic leveling at the source module output is not available with the
OEM Source Module selection.
2.
Press
Frequency
>
(3 of 3)
>
Source Module
. Toggle the
Keysight 8355x
Source Module Off On
softkey to Off.
3.
Toggle the
OEM Source Off On
softkey to On.
4.
Press
OEM Source Module Config
>
Standard WR Freq Bands
>
WR8
90-140GHz.
The selections in the Standard WR Freq Bands menu are pre-defined
frequency ranges and multipliers for the most common external source
module frequency ranges. They are provided for setup convenience. If your
source module has a frequency range not listed in the list of pre-defined
setups, use the
Min Band Freq
,
Max Band Freq
, and
Freq Multiplier
softkeys to manually set the range and PSG frequency display multiplier.
For more information on these manual settings, refer to the
E8257D/67D
PSG Signal Generators Key Reference
.
5.
If the
RF OFF
annunciator is displayed, press
RF On/Off
.
To obtain flatness-corrected power, refer to