Keysight E8357D/67D & E8663D PSG User’s Guide 75
Basic Operation
Using Data Storage Functions
This deletes all registers in the selected sequence.
Deleting All Sequences
CAUTION
Be sure you want to delete the contents of
all
registers and
all
sequences in
the instrument state register.
1.
Press
Preset
.
2.
Press the
Recall
or
Save
hardkey.
Notice that the
Select Seq
softkey shows the last sequence that you used.
3.
Press
Delete All Sequences
.
This deletes all of the sequences saved in the instrument state register.
8757 Network Analyzer Save and Recall Functions
The 8757 network analyzer family can save and recall signal generator
instrument states although communication between the instruments is limited.
A clear register command from the 8757 will cause the signal generator to
replace a register’s contents with default values. Default values can be cleared
from the signal generator by using the
Delete All
softkey menu or by using the
corresponding SCPI (Standard Commands for Programmable Instruments)
command.
The signal generator does not communicate directly with the 8757 network
analyzer. If the 8757 saves an instrument state to a signal generator register
and the user deletes that register, the 8757 will not recognize the deletion. An
attempt, by the 8757, to recall a deleted state will cause the PSG to generate
the error message: +700 “State Save Recall Error...”.