Keysight E8357D/67D & E8663D PSG User’s Guide 117
Basic Digital Operation
Using Waveform Markers
1. Clearing Marker Points from a Waveform Segment
When you set marker points they do not replace points that already exist, but
are set
in addition
to existing points. Because markers are cumulative, before
you set points, view the segment (
) and remove any unwanted points.
With all markers cleared, the level of the event output signal is 0V.
Clearing All Marker Points
1.
In the Marker Utilities menu (
), press
Set Markers
.
2.
Highlight the desired waveform segment.
In an ARB format, there is only one file (
AUTOGEN_WAVEFORM
) and it is
already highlighted.
3.
Highlight the desired marker number:
Press
Marker 1 2 3 4
.
4.
For the selected marker number, remove all marker points in the selected
segment:
Press
Set Marker Off All Points
.
The Set Marker display
These softkeys
change the range
of waveform
sample points
shown on the
marker display.
Each press of the
softkey changes
the sample range
by approximately
a factor of two
Marker
points on
first sample
point
First sample
point shown on
display
Select a segment
The display below shows the I and Q components of the waveform, and
the marker points set in a factory–supplied segment.