Keysight ZA0060A Quick Start Guide
50
Test Steps and Description
Signaling PER
This step must be a child step of
. This test step acquires the WLAN signaling PER
after the WLAN link is established.
Step Settings
– Repeat: Set the number of PER readings to be acquired within the test.
– Downlink (DL) Power: Select the DL power calculated depending on the Systems Offset.
– System Offsets: Set system offset for channels. If the selected channel is not listed in the System
Offsets, the software will interpolate the offsets using the two nearest channels. For example, the
software will interpolate the offsets of the selected channel 8 using the values from channel 6 and 11.
– PER limit: Set the upper limit of PER. Results that exceed this PER limit will be reported as Fail.
– Packet Count: Set number of packets sent to the DUT from 34999B module to calculate PER.
– Packet Size: Set the size of the transmitted data packet size from 24 to 1400 bytes.
– Packet Interval: Set the interval time (in milliseconds) between data packets.
– Timeout: Set the timeout to receive each packet response during the PER test. Range from 100 to
2000 milliseconds.