Keysight ZA0060A Quick Start Guide
43
Bluetooth
®
5 Signaling PER Sensitivity Search Test Plan
To validate the sensitivity of your DUTs, add one of the test steps below in your
test plan. As the arrangement of these steps is important, some of the test steps
are disabled at different positions to avoid the creation of invalid test plans.
The power values set must be divisible by 0.5 and be between -100 to -40 dBm.
Test Steps and Description
Linear Adjustment
Performs PER test by sweeping the output powers from high to low or from low to high.
Advanced Settings:
– The PHY, PER Method, Response Iteration, Early Fail Count, and Reconnect Power parameters are as
described under
test step.
– Step size: Select a sweep step size of 1 or 0.5 dB.
– Direction: A high to low direction will sweep the power levels from maximum to minimum. The
opposite applies when selecting a low to high direction.
– Delay between step: Set the delay time (in seconds) between steps.
– Preferred Mode: Select Active Scan or Connect Request. Selecting Connect Request will require
additional selection of PHY, PER Method, and Response Iteration in the Connect Request tab.
– Reconnect Retries: Sets the number of times to reestablish connection to the DUT.
– Scan Delay from Measure Advertising Interval Step: Select the check box to use the Interval settings
test step or clear the check box to enter the desired value in the
Custom Scan Delay option (default value is set at 170 milliseconds).